Calibrating microbiological strains is one of the most technical procedures performed in a microbiology laboratory.

These quantified strains are used for growth promotion testing of culture media, challenge tests, method validation, and even positive controls.

Today, many laboratories rely on ready-to-use calibrated strains. However, some microbiologists still prepare, store, and calibrate their own strains in-house.

How do they do it in practice?

What are the advantages and limitations of this approach?

To answer these questions, we spoke with Karine Fréry, who shares her hands-on experience with the preparation, storage, and calibration of microbiological strains in the laboratory.


Discover the Top 4 QC Strains for Pharma

We asked the SuperMicrobiologists in a big survey.
Here’s what they said.

What Are In-House Quantitated QC Strains Used For?

Karine, could you introduce yourself?

My name is Karine Fréry, and I have been working as a microbiologist in French-speaking Switzerland for the past 24 years.

For 23 years, I managed a Quality Control laboratory providing microbiological testing services for the environmental, food, cosmetics, and pharmaceutical industries.

For the past year, I have been working as the Laboratory Manager of the Microbiology Department within a medical testing laboratory.


For which applications do you use in-house calibrated strains?

Ah, strains… the treasure of microbiologists!

I mainly used in-house calibrated strains in Quality Control (QC) laboratories for quantitative microbiological analyses. These analyses require precise specifications and allow conclusions regarding compliance or non-compliance against limits defined in standards and regulations.

For example, we used them for:

Shelf-life and preservation studies in the food industry to compare preservatives or physical preservation processes such as heat treatment, irradiation, or modified atmosphere packaging.

Growth Promotion Testing (GPT) of culture media with a target of 10 to 100 CFU per 90 mm plate. For example, on R2A agar for the microbial enumeration of bulk purified water according to European Pharmacopoeia monograph 0008, using Pseudomonas aeruginosa as the growth control organism.

Growth promotion testing for the quality control of non-sterile products according to European Pharmacopoeia chapters 2.6.12 and 2.6.13 using Aspergillus brasiliensis, Bacillus subtilis, Candida albicans, Pseudomonas aeruginosa, Staphylococcus aureus, Escherichia coli, and Salmonella Abony.

Preservative efficacy (challenge) testing in cosmetics according to ISO 11930 or European Pharmacopoeia chapter 5.1.3.

Cold-chain growth studies using mixed cultures of Listeria monocytogenes according to ISO 20976-1.

Biocide efficacy testing according to EN 13397, EN 1276, and EN 1650.


Discover the Top 4 QC Strains for Food

We asked the SuperMicrobiologists in a big survey.
Here’s what they said.

Where Do Microbiological QC Strains Come From?

Where do microbiological strains come from, and in what format are they usually supplied?

The strains used in laboratories generally come from reference culture collections such as ATCC, NCTC, DSMZ, or CIP, depending on the country.

They are often supplied between the second and fourth passage. The further the strain is from the original parent culture, the lower the cost.

They are usually supplied in a dehydrated form either on a swab that can be streaked directly onto an agar plate, or as pellets (lenticules) that need to be rehydrated before inoculation.

Their shelf life is typically between 18 months and 2 years after production. Suppliers generally avoid selling strains with a short remaining shelf life.

Depending on the strain, storage conditions may be -20°C, -80°C, or between 2°C and 8°C.


How many passages are allowed?

The strain bank should allow the preparation of working cultures originating from no more than a fifth passage.

The goal is to preserve the original properties of the parent strain, including its molecular, biochemical, and morphological characteristics.


By the way, what exactly is a “passage”?

A passage corresponds to a subculture made from a reference strain. It is the creation of a new generation through cell division from the parent culture. Each daughter cell can then give rise to another generation.

In practice, after multiple generations, the original characteristics of the strain can no longer be guaranteed.


“For strain banking, we used sterilized milk prepared from autoclaved powdered milk.“

Preparing Stock Cultures and Creating a Strain Bank

How do you prepare stock cultures?

The strain is first recovered according to the supplier’s instructions, with or without a rehydration step, using the appropriate culture medium and incubation conditions.

You need to be very careful. Poor handling can either kill the strain or contaminate it.

Recovering the strain automatically creates an additional passage.

For example, if you purchase a third-passage strain, the recovered culture becomes the fourth passage, which is then frozen.

In our laboratory, we inoculated three agar plates from the original swab. We then spread the inoculum with a sterile loop before incubating under optimal conditions.

Once the culture was well developed, we performed purity and identification testing before banking the strain. This could include biochemical testing, MALDI-TOF analysis, or Sanger sequencing.

Of course, this allowed us to verify the identity of the purchased strain, but it also helped us check our own laboratory practices and identification methods. In a way, it was an internal quality control exercise.


How do you create the strain bank?

For strain banking, we used sterilized milk prepared from autoclaved powdered milk.

Cryobead systems stored in glycerol remain the gold standard, but milk works very well and costs much less.

The lipids present in milk effectively protect cell membranes during freezing.

We added 10 mL of sterile milk to each agar plate and quickly scraped the surface to prevent the liquid from being absorbed by the agar.

The harvest from all three plates was then collected using sterile disposable transfer pipettes into a 50 mL tube.

The suspension was carefully aliquoted into sterile microtubes containing 200 µL, 500 µL, or 1 mL.

All manipulations were performed inside a microbiological safety cabinet using sterile disposable equipment to avoid contamination.

In the end, we could easily prepare around one hundred microtubes of a fourth-passage strain for approximately €100.


How were the strains stored afterward?

Each laboratory should define its own storage period based on identification and quantification studies.

In our case, the strains were stored at -25°C because we did not have access to a -80°C freezer.

For the strains used most frequently, we renewed the stock every 12 months.

We had validated a storage period of less than 24 months, although some rare strains were occasionally kept for more than 10 years.

Quantified the Microbiological QC Strains

After freezing, how did you “restart” the strain?

We always thawed the aliquots on Friday so that we could start the tests the following Tuesday. The aliquot was allowed to thaw for about 30 minutes at room temperature inside a microbiological safety cabinet.

We then streaked the suspension onto an agar plate to create a new generation.

On Monday, we checked the purity of the culture. On Tuesday, we proceeded with the quantification of the strain.


How did you quantify the strains?

Beforehand, we had established nephelometric correlations between McFarland values and actual microbial concentrations, taking into account the culture medium, incubation temperature, and incubation time.

We used a Densichek nephelometer.

For example, we knew that a McFarland value of 0.53 corresponded to approximately 128 million cells per mL for Staphylococcus saprophyticus.

This allowed us to obtain a quick estimate of the concentration, up to approximately 8 log CFU/mL.

We then prepared a standard dilution series using sterile physiological saline or sterile saline peptone solution.


How was the calibration performed in practice?

We picked several colonies until we reached the desired McFarland value in a 5 mL NaCl ampoule.

This suspension became our Tube 0.

We then prepared a standard dilution series from this McFarland tube.

Careful serial ten-fold dilutions were performed by transferring 1 mL into 9 mL of either sterile 0.9% saline solution or sterile saline peptone solution.

It was very important not to use a nutrient broth at this stage.

Otherwise, the microorganisms would start multiplying, which would distort the results.

The dilution series was then plated onto agar in order to determine the actual concentration being inoculated.

To do this, we plated 1 mL of the 10⁻⁴ to 10⁻⁷ dilutions onto general-purpose media to accurately determine the concentration present in our original Tube 0, which we kept as a reference.


How long can a calibrated strain be used after calibration?

For bacteria and yeasts, we stored the calibrated suspensions in a refrigerator for a maximum of three days.

Beyond that point, we observed significant mortality because physiological saline is not a nutritive medium.

For moulds such as Penicillium roqueforti or Aspergillus brasiliensis, we used a different approach by adding Tween to recover the hydrophobic spores.

These preparations could be stored in a refrigerator for up to six months.

Advantages and Limitations of In-House Calibrated Strains

What are the advantages of performing calibration in-house?

The first advantage is cost.

Laboratories also gain greater autonomy and become less dependent on supplier lead times.

It also allows laboratories to respond more quickly to customer requests and to work with strains that are not available as ready-to-use products.

Finally, in-house calibration is an excellent internal quality control exercise that helps verify the technical expertise of the laboratory.


What should laboratories pay particular attention to when performing their own calibration?

Rigor is essential. All manipulations should be performed inside a microbiological safety cabinet.

The purity of the strain must always be verified.

Time and temperature conditions must be perfectly reproducible.

Cross-contamination must be avoided, and great care must be taken when performing serial dilutions and homogenization steps.

Plate counts are essential to determine the actual concentration being inoculated.

The time between preparing the calibrated suspension and inoculating the test system must also be validated.

For example, for some cosmetic preservative efficacy tests, we validated a maximum delay of two hours between preparation and inoculation… no longer than that.


What are the limitations of in-house calibration?

Some microorganisms are more difficult to handle than others.

Anaerobes, for example, are very sensitive to homogenization and dilution steps.

Spore-forming microorganisms such as Bacillus species and moulds are also more difficult to quantify accurately.

Distinguishing between spores and vegetative cells may require microscopic examination or specific extraction procedures.

Finally, it is difficult to exceed concentrations of approximately 10⁸ microorganisms per mL in the initial McFarland suspension.

In-House Calibrated Strains or Ready-to-Use Solutions: Which One Should You Choose?

You have also used ready-to-use calibrated strains. So in which situations would you recommend preparing and calibrating strains in-house rather than using a commercial ready-to-use solution?

I would definitely recommend in-house calibrated strains for R&D applications, such as challenge tests, where protocols often need to be adjusted and optimized at the beginning.

They are also very attractive for laboratories that need large quantities of strains every week.

The cost savings can be substantial.

A ready-to-use calibrated strain can cost anywhere from €10 to €500, depending on the supplier and the level of precision required.

An in-house calibrated strain typically costs between €1 and €3, excluding technician time.

Once the calibration curves have been validated, sufficient data have been collected, and the process is fully under control, it is entirely possible to present the method to accreditation assessors or GMP inspectors.

Once approved, in-house calibrated strains can be a very effective alternative to ready-to-use solutions.

They do require more hands-on work, but good laboratory organization often makes it possible to optimize the time involved.

Ready-to-use solutions obviously provide convenience and excellent traceability.

However, preparing your own strains also brings a real sense of technical accomplishment.

Ultimately, the choice between in-house calibrated strains and ready-to-use solutions depends mainly on the laboratory’s needs, resources, and level of expertise.

Cela demande un peu plus de manipulation, mais l’organisation du laboratoire permet souvent d’optimiser ce temps.

Les solutions ready-to-use apportent évidemment du confort et une excellente traçabilité.

Mais préparer ses propres souches apporte aussi une vraie satisfaction technique.

Finalement, le choix entre souches calibrées maison et souches prêtes à l’emploi dépend surtout des besoins, des ressources et du niveau de maîtrise du laboratoire.