What does the new EN 12469 standard change for BSC users?

Published on 30 April 2026 7 min


This article is also available as a podcast in more than 20 languages. Click the ⚙️ Settings icon on YouTube, then choose Audio Track and Subtitles to watch it in your preferred language.



Since November 2025, a new standard has been introduced for Class II Biological Safety Cabinets (BSCs):

EN 12469-2, titled Biological safety cabinets – Part 2: BSC class II

But in practical terms, what will this update change for us, microbiologists who use BSCs every day?

To find out more, we spoke with Cédric Herry, who was part of the working group behind this new standard.


Discover the Top 5 BioSafety Cabinets

We asked the SuperMicrobiologists in a big survey.
Here’s what they said.




Hello Cédric, can you introduce yourself?

I’m Cédric Herry, trained as a chemist and microbiologist. I’m currently VP of Innovation at Erlab, a manufacturer of safety cabinets for chemistry and microbiology.

I’m also the chairman of the French standardization committee UNM 61, which works on this type of equipment. And I’m an expert in the European working group WG8, which specifically worked on microbiological safety cabinets, what we call BSCs.




So this standard only applies to Europe?

For now, yes. It’s not yet an international ISO standard. But discussions are starting at ISO level. So we could imagine that within the next 5 to 10 years, an international BSC standard could emerge.

In that case, Europeans would likely bring proposals from the European standard. Americans would bring their NSF 49. China also has its own standard.

The goal would be to reach an international consensus.


Is EN 12469 a new standard or an update?

It’s an update, but a major one. The first version of the European BSC standard dates back to 2001 and had barely changed since then.

In more than 20 years, a lot has evolved: technologies, user expectations, and the other standards referenced in the document. So it was time for a deep revision.

EN 12469: what are the main changes for BSCs?

What is the first major change?

The first change is the structure.

Before, there was a single EN 12469 document covering Class I, II, and III BSCs.

Now, the standard is split into several parts:

  • NF EN 12469-1: definitions and general requirements

  • NF EN 12469-2: Class II BSCs

  • NF EN 12469-3: Class III BSCs

  • NF EN 12469-4: Class I BSCs

  • NF EN 12469-5: installation tests, routine tests, and maintenance

For now, the published parts mainly concern Class II BSCs, which represent about 98% of the market (parts 1, 2, and 5).


“The new standard is more of a performance standard.”




Has the philosophy changed?

Yes and this is probably the most important change. The old standard was often seen as a design standard. It gave fixed values: a specific air velocity here, another one there.

The new standard is more performance-based.

The idea is simple: what matters is that the BSC protects the operator, the sample, and the environment. The technology or exact air velocities matter less, as long as performance is demonstrated.


So air velocities don’t disappear completely?

No, they are still important. But instead of applying a single fixed value from the standard, we now rely more on the manufacturer’s specifications.

The manufacturer must clearly state under which conditions the BSC ensures protection. This information must appear in the user manual, but also on the BSC itself, via labels, stickers, or a screen.

For metrology, this changes the approach. Instead of checking a generic value, you now verify the values specific to that BSC.

What does this change for manufacturers?

It gives them more flexibility. They can optimize the balance between safety, energy consumption, noise, and user comfort.

Higher air velocities often mean more noise and higher energy use. If a manufacturer can ensure performance with better airflow optimization, it’s a win for the user.

It could also help manufacturers differentiate themselves more.


Does the new standard introduce testing under degraded conditions?

Yes, and this is a very interesting point. The standard includes performance tests under so-called degraded conditions, called “provocation points.”

The idea is to test the BSC with variations in airflow, for example, reduced inflow velocity or changes in downflow.

This ensures that the BSC remains effective over a range of operating conditions, not just at an ideal point.

For users, this provides extra reassurance and safety.

EN 12469: what changes for BSC users?

Do users need to replace their BSC?

No. If your BSC is 6 months, 1 year, or even 12 years old, there is no obligation to replace it.

Your BSC remains compliant with the standard that applied when it was placed on the market. The new standard mainly applies to new Class II BSCs.


What changes for future metrology?

The test equipment doesn’t necessarily change. But how results are interpreted may change.

Before, we checked relatively standardized air velocity values. Now, we need to rely more on the manufacturer’s specifications.

So for new BSCs, metrology must take into account the declared performance and the values indicated on the equipment.


Does the standard provide more information on decontamination?

Yes, and this is very important. Manufacturers must provide clearer instructions on how to decontaminate the internal volume of the BSC.

This includes not only the work surface, but also other areas inside the cabinet. They must also provide guidance on safe filter replacement.


Does the sash opening height change?

No. There is no minimum or maximum height specified in the standard. Again, the focus is on performance.


Does the standard change anything regarding noise?

Yes. The noise measurement protocol is now better harmonized.

Previously, it was difficult to compare values from different manufacturers because measurement conditions varied (microphone position, operating mode, distance, etc.).

The new standard defines these conditions more clearly. For users, this should make it easier to compare noise levels in brochures.


“When buying a new BSC, make sure it complies with the new standard, with particular attention to ‘provocation points’.”




Is there better alignment with GMP (Good Manufacturing Practices)?

Yes, this is another positive point. Previously, it could be difficult to reconcile GMP requirements with the old EN 12469, especially regarding airflow ranges.

The new standard should allow better alignment between GMP requirements and BSC performance requirements. This should make discussions easier between users, manufacturers, quality teams, and metrology experts.


Any final advice for labs buying a new BSC?

The standard places more emphasis on installation, maintenance, documentation, and end-of-life management.

A good BSC is not just something you buy from a catalog. It must be properly installed, qualified, maintained, and safely decommissioned at the end of its life.

When purchasing a new BSC, make sure it complies with the new standard, especially regarding the “provocation points” mentioned earlier.

And if there is any doubt, it can be useful to include a clause in the purchase order: if the BSC fails initial on-site qualification, the supplier must take it back.

A manufacturer confident in their equipment should have no issue with this kind of commitment.


We would like to thank Cédric Herry for answering all our questions.

It’s not that we don’t like reading standards… but it’s definitely easier when someone explains them clearly.


If you have any questions about this new standard, feel free to ask in the comments. We’ll do our best to answer them.





3 responses

    1. Hi Carin,

      Here is Cedric’s answer to your question :

      “The integrity (leak) test is defined in EN ISO 14644-3. It is also referenced in Part 1 of the new EN 12469 as follows:
      6.9.2 Testing
      6.9.2.1 Filter leak test
      The leakage test of the filter system shall be performed according to EN ISO 14644-3.”