Key takeaways
- Between samples, Pauline’s laboratory decontaminates the perforated lids using wipes impregnated with isopropyl alcohol (IPA).
- The IPA must have completely evaporated before the next sampling operation begins.
- The choice of wipes and the decontamination procedures have been validated at the site.
This article is also available as a podcast in more than 20 languages. On YouTube, click the ⚙️ Settings icon, then select Audio track and Subtitles to watch it in your preferred language.
Between two air samples, what do you do with the perforated lid of your microbial air sampler? Do you replace it with a sterilized one, or decontaminate it before reusing it?
In the laboratory of Pauline Hars, microbiology laboratory manager at a sterile injectable drug manufacturing site, the perforated lids are decontaminated between samples. But choosing the right wipes and allowing the alcohol to evaporate completely require particular attention.
She explains how air sampling is organized at her site and the key points she monitors. A practical experience to consider in light of each laboratory’s operating conditions and validation studies.
To start with, how do you perform active air sampling by impaction?
Pauline Hars: We place an agar plate in a microbial air sampler, a piece of equipment programmed and qualified to sample a defined volume of air. In our laboratory, we mainly sample one cubic metre of air. The air is impacted onto the agar, which is then incubated to detect the presence of microorganisms.
We don’t use the same equipment everywhere. In Grades C and D, the air samplers are mobile: they are moved from one area to another and decontaminated between uses. In Grade A, they are fixed devices installed inside the isolator and handled by production operators.
Where do you position the air samplers during sampling?
In Grades C and D, we don’t place them directly on the floor. We use dedicated stands designed for the air samplers.
Inside the isolator, the equipment is fixed in place. The installation height was determined based on the risk assessment.
How do you manage the perforated lids between samples?
Given the number of samples we collect, we decontaminate the perforated lids using wipes impregnated with isopropyl alcohol, or IPA, between samples.
We wait until the alcohol has completely evaporated before starting the next sample. This is something we need to pay particular attention to.
We also use other treatments, including occasional autoclaving or decontamination with a sporicidal agent.
Is the choice of wipe also important?
Yes. You need to pay attention to the type of wipe you use, and we have validated these choices at our site.
We had to test several types of wipes before finding a suitable one, particularly to avoid leaving particles behind. So it’s not just the disinfectant that matters: the wipe used to decontaminate the perforated lid also needs to be considered.
In what order do you move the air samplers between areas?
We start in the cleanest areas and move towards the less clean ones. In our setup, we therefore start in Grade C before moving to Grade D.
What does the operator do while the air sampler is running?
The operator stays busy. They can label their plates or prepare the next samples while keeping an eye on the ongoing sampling operation to check for any abnormalities.
They can also start another active air sampling operation, collect a surface sample, or open a settle plate for passive air sampling. So they don’t just stand around waiting next to the air sampler.
A big thank you to Pauline for sharing her experience and these practical insights into air sampling in classified cleanrooms!
And in your laboratory, how do you manage your air sampler perforated lids between samples? Share your practices and challenges in the comments.