Skip to main content

SuperMicrobiologists

Sievers Eclipse: A New Approach to Endotoxin Testing

Published on 3 November 2026 6 min



This article is also available as a podcast in more than 20 languages. Click the ⚙️ Settings icon on YouTube, then choose Audio Track and Subtitles to watch it in your preferred language.




Preparing a standard curve, adding samples, preparing Positive Product Controls (PPCs)… Bacterial Endotoxin Testing (BET) involves quite a few pipetting steps!

And when you’re working with very small volumes, there’s always a risk of making a mistake.

So, could microfluidic technology make endotoxin testing easier?

We discussed this with Meg Provenzano, who has more than 15 years of experience in endotoxin testing and is now Product Manager for the Sievers BET product line at Veolia.

She introduced us to Sievers Eclipse, a microfluidic system designed to simplify BET testing, and explained its advantages and limitations.




How does the Sievers Eclipse technology work?

Meg Provenzano: We have a pretty novel solution that uses microfluidic technology and centrifugal force to distribute the LAL reagent and samples evenly, in a one-to-one ratio.

The system consists of an instrument, software and a microfluidic disc (the Eclipse microplate).

Each Sievers Eclipse microplate can accommodate up to 21 samples and a five-point standard curve. You can configure the standard curve depending on your needs, for example from 50 to 0.005 EU/mL or from 5 to 0.005 EU /mL (Endotoxin Unit).

The standard curve is run in triplicate, and the Sievers Eclipse microplate also includes negative controls.

What makes it interesting is that the standards and Positive Product Controls (PPCs) are already dried and pre-deposited on the microplate.

One of the biggest challenges with BET is preparing the standard curve. Human error can be a problem, and two technicians may prepare their curves slightly differently, which can affect reaction times.

With microfluidics, we’re trying to minimize that human error as much as possible.


What type of BET reagents can be used with Sievers Eclipse?

The Sievers Eclipse is reagent-agnostic, which means laboratories are not limited to a single reagent supplier.

You can use commercially available kinetic chromogenic reagents, and the system is also compatible with commercially available recombinant cascade reagents (rCR).

We also work with one kinetic turbidometric reagent.

Another important benefit is the amount of reagent required.

The system only uses 1 mL of LAL reagent for up to 21 samples and a five-point standard curve.

For comparison, an equivalent test using a traditional 96-well microplate would require approximately 10 mL.

That’s a significant reduction in LAL reagent consumption, which can also help laboratories working toward sustainability goals and reducing their impact on horseshoe crabs.


What are the main advantages for microbiologists working in the lab?

I’ve performed all kinds of BET tests over the years, including gel-clot and kinetic assays.

And I was actually back in the lab recently, working with traditional 96-well microplates.

For me, one of the biggest advantages is not having to prepare the standard curve every day.

That step alone can take 30 to 45 minutes, because you need to vortex your CSE or RSE and then prepare the dilutions.

Another big advantage is having the PPCs already pre-deposited.

With a traditional LAL assay, you typically add a 10 µL spike to the microplate. With recombinant cascade reagents, it’s usually 5 µL.

And honestly, as I’m getting older, my eyesight isn’t what it used to be!

Being able to see whether I’ve added 5 or 10 µL can be difficult.

Having the PPCs already on the microplate removes that pipetting step.

And the other thing I really like is that you only need to pipette each sample once.

You add 55 µL of sample, and the system distributes it evenly into four optical wells: two for the sample and two for the PPCs.

It’s the same principle for the LAL reagent. You add 1 mL into a single port, and the system distributes it accurately with centrifugal technology.

After performing so many BET tests over the years, I really appreciate having fewer pipetting steps.



“For each sample you only have to pipette one time.”



What are the limitations of this technology?

I guess the biggest limitation is that some laboratories simply don’t have 21 samples to test at the same time.

They might only have eight or ten samples.

In that case, we recommend batching the samples, meaning keeping them until you have enough to run them together.

We’ve found that using the Sievers Eclipse can already be advantageous with around 12 samples.

But of course, if you’re going to hold samples before testing them, you need to make sure that the holding time is suitable.



How can laboratories make sure that endotoxins remain recoverable while samples are being held?

I would recommend performing a hold-time study.

And that’s not specific to this technology. Whenever you’re going to hold samples in a container before testing them, you should evaluate the holding time.

The idea is to store the samples under the planned conditions and test them at different time points.

For example, if you’re planning to hold samples for seven days, you might test them at day 0, 3, 5 and 7.

You could even include a worst-case scenario, such as nine days, in case something unexpected prevents you from testing on time.

There are two things you can do.

You can spike your samples with a known amount of endotoxin and check whether you can still recover it after storage.

At the same time, you can also keep unspiked samples in the same containers to look for any effects caused by the containers themselves.

It’s usually a pretty simple study to perform.


A big thank you to Meg Provenzano for introducing us to the Sievers Eclipse and sharing her experience as both a laboratory technician and a Product Manager.

It’s always interesting to discover technologies designed to make everyday laboratory work easier, especially when they address challenges that microbiologists know all too well!

If you have any questions about endotoxins, feel free to ask them in the comments.

Leave a Reply